4068 CapSureFix
Pacing
4068 CapSureFix
VENTRICULAR PLACEMENT
Information

Device Survival Probability


4068_VEN_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr at 147 mo                
% Survival 99.3 98.7 98.7 98.2 97.8 97.2 95.8 95.0 93.5 93.1 92.4 89.7 89.7                
#   1427 1219 1032 832 685 539 405 300 213 133 77 52 49                

System Longevity Study Results

Number of Leads Enrolled in Study 1799
Cumulative Months of Follow-Up 95888
Number of Leads Active in Study 162
Information 51

Qualifying Complications

Conductor Fracture 3
Extra Cardiac Stimulation 2
Failure to Capture 22
Failure to Sense 3
Impedance Out of Range 11
Insulation (not further defined) 1
Oversensing 7
Unspecified Clinical Failure 2
ATRIAL PLACEMENT
Information

Device Survival Probability


4068_ATR_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr 10 yr 11 yr 12 yr 13 yr at 165 mo              
% Survival 98.9 98.7 98.2 98.0 97.4 97.2 96.7 95.3 94.6 93.7 92.5 91.4 89.9 86.2              
#   1907 1637 1370 1116 897 737 577 472 358 256 189 135 84 43              

System Longevity Study Results

Number of Leads Enrolled in Study 2413
Cumulative Months of Follow-Up 133591
Number of Leads Active in Study 266
Information 77

Qualifying Complications

Conductor Fracture 3
Extra Cardiac Stimulation 2
Failure to Capture 22
Failure to Sense 13
Impedance Out of Range 9
Insulation (ESC) 2
Insulation (MIO) 2
Insulation (not further defined) 2
Lead Dislodgement 8
Oversensing 11
Unspecified Clinical Failure 3
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 5
Conductor Fracture 3
Extra Cardiac Stimulation 1
Failure to Capture 23
Failure to Sense 5
Impedance Out of Range 2
Insulation Breach 1
Lead Dislodgement 31
Unspecified Clinical Failure 4
Information

Distribution Data

US Market Release1996-03-29
CE Approval Date
Registered USA Implants123828
Estimated Active USA Implants30315
US Market Release1996-03-29
CE Approval Date
Estimated WW Distribution187329
Information

USA Returned Product Analysis

Conductor Fracture 50
Insulation Breach 163
Crimp/Weld/Bond 0
Other 95
Information

WW Returned Product Analysis

Conductor Fracture 63
Insulation Breach 168
Crimp/Weld/Bond 1
Other 247
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Active Screw In
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Atrium or Right Ventricle
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LCE
Lead Body Diameter 7.2 FR
Insulation Polyurethane
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of March 12, 2013