4024 CapSure SP
Pacing
4024 CapSure SP
ALL
Information

Device Survival Probability


4024_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr 6 yr 7 yr 8 yr 9 yr at 117 mo
% Survival 99.9 99.8 99.8 99.8 99.8 99.8 99.8 99.8 98.8 98.8
#   870 685 556 416 323 229 167 111 63 52

System Longevity Study Results

Number of Leads Enrolled in Study 1215
Cumulative Months of Follow-Up 50928
Number of Leads Active in Study 13
Information 4

Qualifying Complications

Failure to Capture 3
Insulation (not further defined) 1
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 12
Conductor Fracture 11
Extra Cardiac Stimulation 2
Failure to Capture 103
Failure to Sense 15
Impedance Out of Range 6
Insulation Breach 1
Lead Dislodgement 48
Oversensing 1
Unspecified Clinical Failure 20
Information

Distribution Data

US Market Release1991-10-01
CE Approval Date
Registered USA Implants217472
Estimated Active USA Implants41101
US Market Release1991-10-01
CE Approval Date
Estimated WW Distribution277530
Information

USA Returned Product Analysis

Conductor Fracture 27
Insulation Breach 178
Crimp/Weld/Bond 0
Other 16
Information

WW Returned Product Analysis

Conductor Fracture 31
Insulation Breach 202
Crimp/Weld/Bond 0
Other 49
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Tines
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Right Ventricle
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LAJ
Lead Body Diameter 5.7 FR
Insulation Polyurethane
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050