3830 SelectSecure
Pacing
3830 SelectSecure
VENTRICULAR PLACEMENT
Information

Device Survival Probability


3830_VEN_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr at 69 mo        
% Survival 99.5 99.5 99.5 99.5 99.5 99.5        
#   388 229 142 90 63 49        

System Longevity Study Results

Number of Leads Enrolled in Study 475
Cumulative Months of Follow-Up 15455
Number of Leads Active in Study 290
Information 3

Qualifying Complications

Impedance Out of Range 1
Lead Dislodgement 2
ATRIAL PLACEMENT
Information

Device Survival Probability


3830_ATR_SURV
    1 yr 2 yr 3 yr 4 yr 5 yr at 63 mo        
% Survival 99.2 99.2 99.2 99.2 99.2 99.2        
#   612 337 173 98 63 54        

System Longevity Study Results

Number of Leads Enrolled in Study 769
Cumulative Months of Follow-Up 22098
Number of Leads Active in Study 509
Information 6

Qualifying Complications

Cardiac Perforation 1
Extra Cardiac Stimulation 1
Failure to Capture 1
Failure to Sense 1
Lead Dislodgement 2
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Cardiac Perforation 5
Conductor Fracture 1
Failure to Capture 14
Failure to Sense 1
Insulation Breach 1
Lead Dislodgement 27
Oversensing 3
Unspecified Clinical Failure 2
Information

Distribution Data

US Market Release2005-08-03
CE Approval Date2003-01-31
Registered USA Implants20869
Estimated Active USA Implants15902
US Market Release2005-08-03
CE Approval Date2003-01-31
Estimated WW Distribution38659
Information

USA Returned Product Analysis

Conductor Fracture 4
Insulation Breach 12
Crimp/Weld/Bond 0
Other 4
Information

WW Returned Product Analysis

Conductor Fracture 7
Insulation Breach 16
Crimp/Weld/Bond 0
Other 5
Information

Product Characteristics

Lead Type Pacing
Type and/or Fixation Fixed Screw
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Atrium or Right Ventricle
Pace/Sense Polarity Bipolar
More
Serial Number Prefix LFF
Lead Body Diameter 4.1 FR
Insulation Polyurethane (outer) Silicone (inner)
Pace/Sense Connector Style IS-1 BI
High Voltage Connector n/a
Steroid Yes
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050