3830
SelectSecure
VENTRICULAR PLACEMENT
Device Survival Probability
| 1 yr | 2 yr | 3 yr | 4 yr | 5 yr | at 69 mo | ||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| % | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | 99.5 | |||||
| # | 388 | 229 | 142 | 90 | 63 | 49 |
System Longevity Study Results
| Number of Leads Enrolled in Study | 475 |
|---|---|
| Cumulative Months of Follow-Up | 15455 |
| Number of Leads Active in Study | 290 |
Performance Notes For This Model
US Acute Lead Observations
| Cardiac Perforation | 5 |
|---|---|
| Conductor Fracture | 1 |
| Failure to Capture | 14 |
| Failure to Sense | 1 |
| Insulation Breach | 1 |
| Lead Dislodgement | 27 |
| Oversensing | 3 |
| Unspecified Clinical Failure | 2 |
Product Characteristics
| Lead Type | Pacing |
|---|---|
| Type and/or Fixation | Fixed Screw |
| Lead Function | Pacing/Sensing |
| Lead Placement | Transvenous |
| Lead Tip Location | Atrium or Right Ventricle |
| Pace/Sense Polarity | Bipolar |
| Serial Number Prefix | LFF |
|---|---|
| Lead Body Diameter | 4.1 FR |
| Insulation | Polyurethane (outer) Silicone (inner) |
| Pace/Sense Connector Style | IS-1 BI |
| High Voltage Connector | n/a |
| Steroid | Yes |
Data as of March 12, 2013