2187 Attain LV
Left-Heart Pacing
2187 Attain LV
ALL
Information

Device Survival Probability


2187_SURV
    1 yr 2 yr 3 yr 4 yr at 54 mo          
% Survival 99.1 99.1 99.1 99.1 99.1          
#   109 92 71 57 51          

System Longevity Study Results

Number of Leads Enrolled in Study 134
Cumulative Months of Follow-Up 6667
Number of Leads Active in Study 14
Information 1

Qualifying Complications

Failure to Capture 1
Information

Advisories For This Model

  • There are no advisories for this model.
Information

US Acute Lead Observations

Extra Cardiac Stimulation 1
Failure to Capture 3
Failure to Sense 1
Lead Dislodgement 9
Information

Distribution Data

US Market Release2001-08-28
CE Approval Date
Registered USA Implants11976
Estimated Active USA Implants2729
US Market Release2001-08-28
CE Approval Date
Estimated WW Distribution23879
Information

USA Returned Product Analysis

Conductor Fracture 1
Insulation Breach 0
Crimp/Weld/Bond 0
Other 4
Information

WW Returned Product Analysis

Conductor Fracture 1
Insulation Breach 0
Crimp/Weld/Bond 0
Other 14
Information

Product Characteristics

Lead Type Left-Heart
Type and/or Fixation Distal Continous Curve
Lead Function Pacing/Sensing
Lead Placement Transvenous
Lead Tip Location Cardiac vein
Pace/Sense Polarity Unipolar
More
Serial Number Prefix LEY
Lead Body Diameter 6.2 FR
Insulation Polyurethane (55D)
Pace/Sense Connector Style IS-1 UNI
High Voltage Connector n/a
Steroid None
Hide this content

Data as of March 12, 2013

www.medtronic.com/productperformance 251050