Performance Note: Tests and Observations for Clinical Assessment of Chronic Pacing Leads
Chronic Pacing Leads
Original Date of Communication:
January 1992
View specific models that this applies to
Test/Observation |
Possible Insulation Failure |
Possible Conductor Failure |
Possible Other System Failure |
Effect on Test/ Observation |
||||||||||
Pacing Impedance |
Sudden and Significant |
Sudden and Significant Increase |
|
|||||||||||
|
Sudden and Significant Increase, Especially in Bipolar System |
Sudden and Significant Increase |
|
|||||||||||
Electrograms |
Sudden and Significant Decrease in Amplitudes and/or Slew Rates for P and/or R Waves |
Sudden and Significant Decrease or Disappearance of Amplitudes and/or Slew Rates for P and/or R Waves |
|
|||||||||||
Waveform Analysis |
Sudden Increase in Ratios of Leading-Edge Voltages to Trailing-Edge Voltages (i.e., over 25% increase) |
Intermittent or No Pacer Artifacts (Even in Asynchronous Mode) |
Improper IPG/Lead Connection |
Intermittent or No Pacer Artifacts |
||||||||||
Radiographs |
Not Discernible |
Visual Observation of Conductor/Connector/ Electrode Fracture (Sometimes Discernible) |
Dislodgement or Perforation. |
Sometimes Discernible |
||||||||||
Visual Inspection |
Insulation Breach and/or Degradation, or Ligature Cut-Through |
Not Easily Discernible |
Connector Defect or Connector Pulled Apart. Improper IPG/Lead Connection. |
Sometimes Discernible |
||||||||||
Pectoral Muscle Stimulation |
Sudden Onset, Especially in Bipolar System |
Connector Defect in Bipolar or Unipolar. Hypersensitivity to Unipolar Pulse Generator Can. Anti-Stim Coating or Protection Deficient. |
||||||||||||
Phrenic Nerve/ |
Sudden Onset in Bipolar or Unipolar Systems |
Perforation or Displacement of Atrial Lead (Phrenic Nerve) |
||||||||||||
Pacemaker ECG Stimulus |
Sudden Onset and Significant Change, Especially in Bipolar System (Increase in Size) |
Sudden Changes, Usually a Decrease in Size |
Perforation or Dislodgement. Connector Defect. Improper |
Sometimes Discernible |
||||||||||
Oversensing |
Sudden Onset, Especially in Bipolar Systems |
Physical Contact between the Electrode(s) on the Lead and that of Another Lead. Inappropriate IPG Parameter Setting. Improper IPG/Lead Connection. |
Sometimes Discernible |
|||||||||||
Undersensing |
Sudden Onset in Either Unipolar or Bipolar Systems |
Sudden Onset in Either Unipolar or Bipolar Systems |
Dislodgement or Perforation. Infarct at Electrode Site. Electrolyte Imbalance. Inappropriate IPG Parameter Setting. Improper IPG/Lead Connection. |
Sometimes Discernible |
||||||||||
Loss of Capture |
See “Pacing Thresholds" Above |
See “Pacing Thresholds” Above |
See “Pacing Thresholds” Above |