Performance Notes
Ensuring the Accuracy of Battery Longevity Estimates

Ensuring the Accuracy of Battery Longevity Estimates

View specific models that this applies to

Purpose of This Information

This article is intended to help the clinician understand how Medtronic estimates CRT-D, ICD, and IPG device longevity and Medtronic’s performance against these estimates.

Device Longevity and Battery Depletion

The device service life ends when the usable battery capacity is depleted. The time to battery depletion depends on three factors:

  • The amount of electrical energy expended in providing therapy to the patient
  • The amount of energy consumed by the electronic circuitry to perform the functions of the device (e.g., operating the microprocessor, telemetry, memory, and charging component)
  • The energy capacity of the battery

Medtronic has developed a statistical model for device longevity that accounts for each of these factors, and has validated the model with real time clinical performance. During the development of its products, Medtronic engineers characterize device longevity using this model. Testing begins during development and continues after market release to ensure the accuracy of device longevity estimates.

Using Survival Curves to Assess Longevity

The survival curves in the Product Performance Report represent the composite experience of thousands of devices over a wide range of programming options and patient use conditions. While the curves are useful for understanding the overall performance of a population of devices, they cannot be used to accurately predict the longevity of a specific device in a specific patient. To get a longevity prediction for a specific device, the longevity model must be used. The model is available by contacting Medtronic’s Technical Services Department.

Because the survival curves are an aggregate result, the Reference pages in the Product Performance Report include several longevity estimates for a range of use conditions. These longevity estimates are mean values calculated for the parameters given. This range of longevity estimates can be compared to the survival curve including normal battery depletion to assess the overall clinical performance of a device model against the original longevity estimates.

If most of a device model’s population is being used at nominal parameters and conditions, the time at which the survival curve including normal battery depletion equals 50% should approximate the midpoint in the range of longevity estimates.

If devices tend to be used at conditions that consume more or less energy than nominal, then the time at which the survival curve equals 50% should tend toward the lower or higher end of the range of longevity estimates, respectively.

Figure of the estimated longevity values

Specific Models This Applies To

7088 Preva DR
7089 Preva DR
7107 Minuet
7108 Minuet
7227 Gem VR
7227B GEM
7227CX GEM
7227D GEM
7227E GEM
7230B Marquis VR
7230Cx Marquis VR
7230E Marquis VR
7231Cx GEM III VR
7232B Maximo VR
7232Cx Maximo VR
7232E Maximo VR
7271 GEM DR
7274 Marquis DR
7275 GEM III DR
7277 InSync Marquis
7278 Maximo DR
7279 InSync III Marquis
7285 InSync III Protect
7288 Intrinsic
7289 InSync II Marquis
7290Cx Onyx
7295 InSync II Protect
7297 InSync Sentry
7298 InSync Sentry
7299 InSync Sentry
7303 InSync Maximo
7304 InSync Maximo
7860 Prodigy DR
7861 Prodigy DR
7862 Prodigy DR
7960i Thera-i DR
7961i Thera-i DR
7962i Thera-i DR
8040 InSync
8042 InSync III
8085 Prevail S
8086 Prevail S
8088 Preva SR
8089 Preva SR
8158 Prodigy SR
8160 Prodigy SR
8161 Prodigy SR
8162 Prodigy SR
8330 Minix ST
8331 Minix ST
8331M Minix ST
8340 Minix
8341 Minix
8341M Minix
8342 Minix
8424 Legend II
8426 Legend II
8427 Legend II
8960i Thera-i SR
8961i Thera-i SR
8962i Thera-i SR
8968i Thera-i VDD
A2DR01 Advisa DR MRI
A3DR01 Advisa DR MRI+C82
A5DR01 Advisa DR
ADD01 Adapta D
ADDR01 Adapta DR
ADDR03 Adapta DR
ADDR06 Adapta DR
ADDRL1 Adapta DR
ADDRS1 Adapta DR
ADSR01 Adapta SR
ADSR03 Adapta SR
ADSR06 Adapta SR
ADVDD01 Adapta VDD
AT501 AT500
C10A1 C10 S
C10A2 C10 S
C10A3 C10 S
C10A4 C10 S
C154DWK Concerto CRT-D DR
C164AWK Concerto CRT-D DR AT
C174AWK Concerto CRT-D DR AT
C20A1 C20 SR
C20A2 C20 SR
C20A3 C20 SR
C20A4 C20 SR
C2TR01 Syncra CRT-P
C3TR01 Consulta CRT-P
C4TR01 Consulta CRT-P
C50A1 C50 D
C50A2 C50 D
C50A3 C50 D
C50A4 C50 D
C60A1 C60 DR
C60A2 C60 DR
C60A3 C60 DR
C60A4 C60 DR
C70A3 C70 DR
C70A4 C70 DR
D144DRG Entrust Escudo
D144VRC Entrust Escudo
D153ATG Entrust AT
D153DRG Entrust DR
D153VRC Entrust VR
D154ATG Entrust AT
D154AWG Virtuoso DR
D154DRG EnTrust
D154VRC Entrust VR
D154VWC Virtuoso VR
D164AWG Virtuoso DR
D164VWC Virtuoso VR
D204DRM Secura DR M4
D204TRM Consulta CRT-D M4
D204VRM Secura VR M4
D214DRM Secura DR
D214TRM Consulta CRT-D
D214VRM Secura VR
D224DRG Secura DR
D224TRK Consulta CRT-D
D224VRC Secura VR
D234DRG Secura DR
D234TRK Consulta CRT-D
D234VRC Secura VR
D264DRM Maximo II DR
D264TRM Maximo II CRT-D
D264VRM Maximo II VR
D274DRG Virtuoso II DR
D274TRK Concerto II CRT-D
D274VRC Virtuoso II VR
D284DRG Maximo II DR
D284TRK Maximo II CRT-D
D284VRC Maximo II VR
D294DRG Virtuoso II DR
D294TRK Concerto II CRT-D
D294VRC Virtuoso II VR
D314DRG Protecta XT DR
D314DRM Protecta XT DR M4
D314TRG Protecta XT CRT-D
D314TRM Protecta XT CRT-D M4
D314VRG Protecta XT VR
D314VRM Protecta XT VR M4
D334DRG Protecta DR
D334DRM Protecta DR M4
D334TRG Protecta CRT-D
D334TRM Protecta CRT-D M4
D334VRG Protecta VR
D334VRM Protecta VR M4
D354DRG Protecta XT DR
D354DRM Protecta XT DR M4
D354TRG Protecta XT CRT-D
D354TRM Protecta XT CRT-D M4
D354VRC Protecta XT VR
D354VRG Protecta XT VR
D354VRM Protecta XT VR M4
D364DRG Protecta DR
D364DRM Protecta DR M4
D364TRG Protecta CRT-D
D364TRK Protecta CRT-D
D364TRM Protecta CRT-D M4
D364VRC Protecta VR
D364VRG Protecta VR
D364VRM Protecta VR M4
D384DRG Cardia DR
D384DRM Cardia DR M4
D384TRG Cardia CRT-D
D384TRM Cardia CRT-D M4
D384VRG Cardia VR
D384VRM Cardia VR M4
D394DRG Egida DR
D394DRM Egida DR M4
D394TRG Egida CRT-D
D394TRM Egida CRT-D M4
D394VRG Egida VR
D394VRM Egida VR M4
DTBX1QQ BLACKWELL
DTBX2QQ BLACKWELL
E10A1 E10 S
E1DR01 EnPulse DR
E1DR03 EnPulse DR
E1DR06 EnPulse DR
E1DR21 EnPulse DR
E2D01 EnPulse 2
E2D03 EnPulse 2
E2DR01 EnPulse 2 DR
E2DR03 EnPulse 2 DR
E2DR06 EnPulse 2 DR
E2DR21 EnPulse 2 DR
E2DR31 EnPulse 2 DR
E2DR33 EnPulse 2 DR
E2SR01 EnPulse 2 SR
E2SR03 EnPulse 2 SR
E2SR06 EnPulse 2 SR
E2VDD01 EnPulse 2 VDD
E50A1 E50 D
E60A1 E60 DR
EMDR01 EnRhythm MRI
EMDR02 EnRhythm MRI
EN1DR01 Ensura MRI
G20A1 G20 SR
G70A1 G70 DR
KD700 Kappa 700 DR
KD701 Kappa 700 DR
KD703 Kappa 700 DR
KD706 Kappa 700 DR
KD901 Kappa 900 D
KD903 Kappa 900 D
KD906 Kappa 900 D
KDR401 Kappa 400 DR
KDR403 Kappa 400 DR
KDR600 Kappa 600 DR
KDR601 Kappa 600 DR
KDR603 Kappa 600 DR
KDR606 Kappa 600 DR
KDR651 Kappa 600 DR
KDR653 Kappa 600 DR
KDR700 Kappa 700 DR
KDR701 Kappa 700 DR
KDR703 Kappa 700 DR
KDR706 Kappa 700 DR
KDR720 Kappa 700 DR
KDR721 Kappa 700 DR
KDR730 Kappa 700 DR
KDR731 Kappa 700 DR
KDR733 Kappa 700 DR
KDR801 Kappa 800 DR
KDR803 Kappa 800 DR
KDR806 Kappa 800 DR
KDR901 Kappa 900 DR
KDR903 Kappa 900 DR
KDR906 Kappa 900 DR
KDR921 Kappa 900 DR
KDR931 Kappa 900 DR
KDR933 Kappa 900 DR
KSR401 Kappa 400 SR
KSR403 Kappa 400 SR
KSR700 Kappa 700 SR
KSR701 Kappa 700 SR
KSR703 Kappa 700 SR
KSR706 Kappa 700 SR
KSR901 Kappa 900 SR
KSR903 Kappa 900 SR
KSR906 Kappa 900 SR
KVDD700 Kappa 700 VDD
KVDD701 Kappa 700 VDD
KVDD901 Kappa 900 VDD
P1501DR EnRhythm DR
RED01 Relia D
REDR01 Relia DR
RES01 Relia S
RESR01 Relia SR
REVDD01 Relia VDD
RVDR01 Revo MRI SureScan
SD200 Sigma
SD203 Sigma 200 D
SD300 Sigma
SD303 Sigma 300 D
SDR200 Sigma
SDR203 Sigma 200 DR
SDR300 Sigma
SDR303 Sigma 300 DR
SDR306 Sigma 300 DR
SED01 Sensia D
SEDR01 Sensia DR
SEDRL1 Sensia DR
SES01 Sensia S
SESR01 Sensia SR
SR353 Sigma
SS100 Sigma
SS103 Sigma 100 S
SS106 Sigma 100 S
SS200 Sigma
SS203 Sigma 200 S
SS300 Sigma
SS303 Sigma 300 S
SSR200 Sigma
SSR203 Sigma 200 SR
SSR300 Sigma
SSR303 Sigma 300 SR
SSR306 Sigma 300 SR
SVDD303 Sigma 300 VDD
SVVI103 Sigma 100 VVI
T20A1 T20 S
T20A2 T20 S
T60A1 T60 DR
T60A2 T60 DR
T70A1 T70 DR
T70A2 T70 DR
VEDR01 Versa DR
www.medtronic.com/productperformance 251050